Patient Setups
These settings are provided as a guideline for initial settings.
Individual patients may require further adjustments based on
their own conditions. Existing protocols within your facility
should always supersede these baseline recommendations.
Please reference the TiControl guide on page 24
for correlating Ti Max/Ti Min settings appropriate
for each disease state and adjust based on
patient’s resting respiratory rate.
Patients with obstructive lung disease have chronic airflow
limitation. These patients have particular difficulty exhaling
air, which leads to air trapping and hyperinflation. These
patients require a longer exhalation, which often leads to
asynchrony with standard bilevel settings.
The recommended settings use a faster rise time to ensure
that the lungs are filled quickly, and a high cycle sensitivity
to provide an earlier cycle to exhalation. The rapid inhalation
and prolonged exhalation will help to prevent auto-PEEP
and preserve synchrony.
Obstructive lung disease
Patients with restrictive lung disease have a difficult
time maintaining the inhalation phase long enough to
ensure adequate tidal volume and gas exchange. This
can be caused by a physical restriction of the lungs or
by neuromuscular weakness.
The recommended settings use a low cycle sensitivity and
a longer Ti Min time to provide a longer inhalation time to
help increase tidal volume and gas exchange.
Restrictive lung disease
Recommended Settings
Obstructive
Lung Disease
IPAP [cm H
2
O] 13
EPAP [cm H
2
O] 5
Ti Max [sec]
1
1.0
Ti Min [sec]
1
0.3
Rise time [ms]
2
150
Trigger sensitivity Medium
Cycle sensitivity High
PS [cm H
2
O] 8
Recommended Settings
Restrictive
Lung Disease
IPAP [cm H
2
O] 11
EPAP [cm H
2
O] 5
Ti Max [sec]
1
1.5
Ti Min [sec]
1
0.8
Rise time [ms]
2
300
Trigger sensitivity High
Cycle sensitivity Low
PS [cm H
2
O] 6
22 S9 VPAP Tx Lab System | Sleep Lab Titration Guide
ResMed.com
VPAP Technology