U.S. Range PS-4-20 Especificações Página 41

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 52
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 40
iVAPS Therapy Titration Protocol
Indications for iVAPS therapy
•Neuromuscular/restrictive disorders
•COPD
•Obesity hypoventilation
Once a patient is fitted with an
appropriate mask, select the
appropriate mask setting.
Evaluate if Target Pt Rate is adequate:
•If central events persist, increase Target Pt Rate by
1–2 BPM every 20 min as needed
For obstructive apneas:
•Increase EPAP by ≥ 1 cm H
2
O every ≥ 5 min to
eliminate obstructive apneas, hypopneas, snoring
and flow limitation
For SpO
2
< 90% with all
respiratory events eliminated:
•Increase Target Va by 0.3 every ≥ 5 min until
desaturations are resolved
Initial iVAPS settings:
•Set Patient Height (eg, 70 inches for 5’10”)
•Set Target Pt Rate equivalent to patient’s
spontaneous respiratory rate (eg, 18 bpm)
•Set Target Va such that Vt is equal to 6ml/kg IBW
•EPAP = 5 cm H
2
O
•Min PS = 4 cm H
2
O
•Max PS = 20 cm H
2
O
Evaluate and titrate:
•Based on Target Pt Rate, Target Va, Vt, SpO
2
and CO
2
compared to baseline
Lab Support 24/ 7 (888) 288-6738 | Sleep Lab Titration Guide 39
iVAPS Titration
Vista de página 40
1 2 ... 36 37 38 39 40 41 42 43 44 45 46 ... 51 52

Comentários a estes Manuais

Sem comentários